Manometer Intrinsically Safe 475-000-FM

Intrinsically Safe handheld digital manometer, range 0-1.000″ w.c. (.2491 kPa), max. pressure 5 psig

Returns: 30-day returns. Buyer pays for return postage.  Please read our terms and conditions

Downloads

Description

Intrinsically Safe Manometer

The Series 475 Intrinsically Safe Handheld Digital Manometer measures positive,
negative, or differential pressures of air and natural gases in ranges from 1 in w.c.
(0.249 kPa) to 150 psi (10.34 bar). The dual push pads on the front panel control the
on/off, auto zero, and pressure unit selection, allowing for simple operation with no set
up needed. When used with a Dwyer® Pitot tube (u), the Series 475 can also be used
as an air velocity gauge.


FEATURES/BENEFITS

• Rugged aluminium case protects the instrument from damage during transport/use
• Large, easy-to-read LCD and simple operation
• FM approved to be intrinsically safe in hazardous locations, Class 1, Div 2, Groups
A, B, C, D, T4 Ta = 70°C


APPLICATIONS

• Monitoring natural gas pressures on boilers and other combustion equipment
• Air velocity monitoring, when used with a Dwyer® Pitot tube (u) and AV calculator
• Field calibration of other instruments
• Monitoring or troubleshooting HVAC systems


SPECIFICATIONS

  • Service: Air and compatible combustible gases.
  • Wetted Materials: Consult factory.
  • Accuracy: ±0.5% FS, 60 to 78°F (15.6 to 25.6°C); ±1.5% FS from 32 to 60°F and 78 to 104°F (0 to 15.6°C and 25.6 to 40°C).
  • Pressure Hysteresis: ±0.1% FS.
  • Temperature Limits: 0 to 140°F (-17.8 to 60°C).
  • Compensated Temperature Limits: 32 to 104°F (0 to 40°C).
  • Storage Temperature Limits: -4 to 176°F (-20 to 80°C).
  • Display: 0.42˝ (10.6 mm) 4 digit LCD.
  • Resolution: See chart.
  • Power Requirements: 9 V alkaline battery, installed non-functional, userreplaceable.
  • Weight: 10.8 oz (306 g).
  • Process Connections: Two barbed connections for use with 1/8˝ (3.18 mm) or
    3/16˝ (4.76 mm) ID tubing. Two compression fittings for use with 1/8˝ (3.18 mm) ID
    x 1/4˝ (6.35 mm) OD tubing for 475-7 & 475-8 only.
  • Agency Approvals: CE, FM approved to Class I, Div 2, Groups A, B, C, D, T4 Ta
    = 70°C.

Our other HVAC products and Sensors – HVAC Products

Have a Question?

If you have any questions, need personalized recommendations, or want to share your thoughts, our dedicated team is here for you. Click the button below to head over to our Contact Us page – your gateway to seamless assistance and exceptional service.